Orbital MRI Protocol: 10 Critical Steps & Fixes

Master the orbital MRI protocol: coronal/axial T1–T2 fat suppression, STIR vs Dixon fixes, optic nerve pathology, and SAR-safe scanning steps.

Orbital MRI Protocol: 10 Critical Steps for Optic Nerve Imaging

⏱ 26 min read Category: Neuroradiology / MRI Protocols ✓ Medically Reviewed Day 6 of the Protocol Mastery Series

At a glance

Sequences used
Coronal & axial T2-weighted TSE, coronal & axial T1-weighted fat-suppressed (pre- and post-contrast), STIR or Dixon fat-suppressed T2, and thin-section high-resolution imaging of the optic nerve–sheath complex with a matrix ≥ 320.
Contrast protocol
10–15 mL (0.1 mmol/kg) gadolinium-based contrast at 1.5 mL/s, followed by a 100 mL saline chaser at 1.5 mL/s.
Artefact reduction
Replacing failure-prone spectral (CHESS) fat saturation with STIR (field-inhomogeneity-insensitive) or two-/three-point Dixon water–fat separation.
Primary pitfall
Local fat suppression failure at the orbital apex and globe periphery, mimicking or masking true pathology.

Introduction

The orbital MRI protocol is one of the most technically demanding studies in neuroradiology because the orbit packs bone, air, fat, muscle, nerve, and globe fluid into a space only a few centimetres across. Every one of those tissues has a wildly different magnetic susceptibility, so the very anatomy that makes the orbit clinically important — the optic nerve, the extraocular muscles, the lacrimal gland — sits at the mercy of local field inhomogeneity. Getting the orbital MRI protocol right means controlling fat suppression, spatial resolution, and motion in a region the size of a walnut, twice, symmetrically, every single time.

This article gives radiographers a reproducible ten-step acquisition sequence, gives radiologists a relaxation-value and pathology reference they can use at the console, and gives referring physicians a plain-language account of why the orbit is unforgiving of shortcuts. We will walk through anatomy, tissue relaxation values, scanning technique, the contrast protocol, SAR management, the top ten pathologies encountered in orbital imaging, and — because this is where most errors actually occur — three full pitfall tables split by professional role, followed by a side-by-side comparison.

Clinical context Orbital and optic-nerve MRI is requested for unexplained visual loss, proptosis, painful eye movement, suspected optic neuritis, orbital trauma, and staging of lacrimal or globe malignancy. Because the differential spans inflammatory, vascular, neoplastic, and infectious disease, the protocol must be sensitive enough to detect a 2 mm optic nerve lesion while still covering the entire orbit, cavernous sinus, and adjacent brain in a single, patient-tolerable acquisition.

Anatomy

The bony orbit is a quadrilateral pyramid formed by seven bones — the frontal, zygomatic, maxillary, lacrimal, ethmoid, sphenoid, and palatine — converging posteriorly at the optic canal and superior orbital fissure. The globe sits anteriorly within intraconal and extraconal fat, itself subdivided by the muscle cone: the four rectus muscles (superior, inferior, medial, lateral) and the superior and inferior oblique muscles, all converging at the annulus of Zinn. The optic nerve — cranial nerve II — runs from the posterior globe through this intraconal fat, through the optic canal, to the optic chiasm, and is itself surrounded by a dural sheath continuous with the intracranial subarachnoid space.

Relevant clinical sub-anatomy

The lacrimal gland occupies the superolateral extraconal compartment and is a frequent site of inflammatory and lymphoproliferative disease. The superior ophthalmic vein drains into the cavernous sinus and dilates in carotid-cavernous fistula. The orbital septum divides pre-septal from post-septal (orbital) space, a distinction that changes the urgency and differential of cellulitis dramatically. The optic nerve–sheath complex has three critical checkpoints for pathology: the intraocular segment (papilloedema), the intraorbital segment (optic neuritis, glioma), and the intracanalicular segment (compressive lesions at the optic canal, the tightest anatomical bottleneck along its course).

MR tissue relaxation values

Understanding baseline relaxation behaviour of orbital tissues at both field strengths is essential for recognising when fat suppression has failed versus when a genuine lesion is present. The table below gives approximate values used for protocol optimisation and differential reasoning.

TissueT1 (ms) 1.5TT1 (ms) 3TT2 (ms) 1.5TT2 (ms) 3TSignal notes
Vitreous humour~1800–2000~2100~1600–1800~1700Long T1/T2 — dark T1, bright T2, mimics CSF
Lens (cortex)~200–300~230~50–60~45Short T2 gives characteristic low signal core
Optic nerve~750–850~900~70–90~75Intermediate; enhances mildly, avidly in neuritis
Extraocular muscle~800–900~950~40–50~45Enlarges/high T2 in thyroid eye disease
Orbital fat~200–250~230~60–80~65Bright T1 — must be suppressed for lesion conspicuity
Lacrimal gland~700–800~850~45–55~50Slightly longer T1 than muscle; enhances homogeneously
Sclera/dura (optic sheath)~500–600~600~20–30~25Low signal on all sequences; thin, easily obscured

These values explain why the orbit is uniquely reliant on fat suppression: orbital fat’s short T1 makes it bright on every routine T1-weighted image, and its intermediate-to-long T2 keeps it bright on standard T2-weighted images too, so without robust suppression it obscures both the optic nerve and any subtle enhancing lesion sitting immediately adjacent to it.

Scanning technique

10-step acquisition workflow

  1. Coil and positioning: dedicated head/neurovascular coil, chin slightly tucked, canthomeatal line perpendicular to the table, both orbits centred symmetrically in the coil.
  2. Localiser: 3-plane scout to confirm symmetry and plan angulation parallel to the optic nerve axis.
  3. Axial T2 TSE through the orbits and brain, matrix ≥ 320, slice thickness ≤ 3 mm, small FOV (16–18 cm).
  4. Coronal T2 fat-suppressed (STIR or Dixon) — the single most diagnostic plane for extraocular muscle and optic nerve assessment.
  5. Axial T1 TSE (non-fat-suppressed) for baseline anatomy and fat/haemorrhage characterisation.
  6. Coronal T1 fat-suppressed pre-contrast, matching the post-contrast plane for direct comparison.
  7. Optic nerve-dedicated thin-slice sequence (2–3 mm, no gap) angled along the nerve for the intracanalicular segment.
  8. Contrast administration per the protocol in the next section.
  9. Post-contrast axial and coronal T1 fat-suppressed sequences, identical geometry to the pre-contrast acquisition.
  10. Quality-control review at the console: confirm symmetric fat suppression bilaterally, adequate SNR, and no motion before releasing the patient.

Scanner comparison: 1.5T vs 3.0T

Parameter1.5T3.0T
Matrix320 × 256384 × 320 (higher achievable resolution)
Fat suppression of choiceSTIR (robust, less field-sensitive)Dixon preferred (higher SNR headroom offsets sequence length)
Slice thickness3 mm2–2.5 mm
SAR headroomGreater; fewer duty-cycle limitsReduced; requires flip-angle/TR management (see SAR section)
Susceptibility artefact at globe/sinus interfaceMilderMore pronounced — shim volume must be tightly confined to the orbit
Typical scan time (full protocol)18–22 min14–18 min (with parallel imaging)
🧠

Standardise anatomy across every orbital case

See how SATPro positioning support helps radiographers reproduce symmetric coil placement scan after scan.

Explore SATPro →

Contrast media protocol

Contrast-enhanced imaging is indicated whenever inflammatory, neoplastic, or infectious orbital disease is suspected, and is essentially mandatory in suspected optic neuritis, orbital cellulitis with abscess concern, and any orbital mass. The standard orbital MRI protocol contrast regimen uses a weight-based gadolinium-based contrast agent (GBCA) at 0.1 mmol/kg, typically 10–15 mL for an average adult, injected at 1.5 mL/s — deliberately slower than neuro-stroke or dynamic pituitary protocols, since the orbit does not require ultra-fast bolus timing. This is followed by a 100 mL saline chaser also at 1.5 mL/s to clear the line and ensure full agent delivery.

Post-contrast imaging must be acquired with fat suppression in both planes; without it, enhancing lesions blend into the naturally bright T1 signal of orbital fat and are easily missed. Renal function (eGFR) should be checked per institutional GBCA safety policy before administration, particularly in patients with known renal impairment.

Safety check Confirm no contraindication to GBCA administration, verify recent renal function where indicated by local policy, and screen for prior contrast reactions before every injection. Pregnancy status should be confirmed per departmental protocol, since GBCAs cross the placenta.

Specific absorption rate

Orbital protocols rely heavily on fast spin-echo (TSE) sequences with multiple closely spaced refocusing pulses, which drives SAR upward — a particular concern at 3T where RF power deposition scales with the square of field strength. SAR limits are governed by IEC 60601-2-33 and align with guidance from the International Commission on Radiological Protection (ICRP), the American Association of Physicists in Medicine (AAPM), and the European Commission Radiation Protection 185 (EC RP 185) series on MR safety.

Operating modeWhole-body SAR limitHead SAR limit
Normal operating mode2 W/kg3.2 W/kg
First-level controlled mode4 W/kg3.2 W/kg (monitored)

Five SAR/dose reduction strategies

  1. Increase TR moderately on T2 TSE sequences to allow more time between RF pulse trains, directly reducing average power deposition.
  2. Reduce echo train length (ETL) or refocusing flip angle (variable flip-angle TSE) to cut per-slice RF energy without proportionally lengthening scan time.
  3. Use parallel imaging acceleration to reduce the number of RF excitations required per unit time.
  4. Favour STIR over multiple CHESS-based fat-sat repeats where duty cycle is tight, since STIR’s inversion pulse can be tuned into an efficient duty cycle compared with repeated spectral saturation pulses.
  5. Confine the imaging volume tightly to the orbit rather than a whole-head FOV, reducing the total tissue volume subjected to RF exposure per sequence.

Top 10 pathologies

1

Optic neuritis

T1: iso/hypo · T2: hyperintense nerve with enhancement
Enlarged, enhancing optic nerve on fat-suppressed T1; protocol impact: fat suppression and post-contrast imaging are essential for detection — without it, enhancement is invisible against orbital fat.
2

Thyroid eye disease

T1: iso · T2: hyperintense enlarged muscle bellies
Fusiform enlargement of extraocular muscles sparing tendinous insertions; coronal T2 fat-sat is the single most diagnostic plane.
3

Optic nerve glioma

T1: hypo/iso · T2: hyperintense, fusiform nerve enlargement
Requires thin-slice dedicated nerve imaging through the intracanalicular segment to assess extension.
4

Optic nerve sheath meningioma

T1: iso · T2: iso/hypo with “tram-track” enhancement
Post-contrast fat-suppressed axial imaging shows the classic tram-track sign; non-fat-sat images can obscure this.
5

Orbital cellulitis / abscess

T1: hypo · T2: hyperintense with rim enhancement
Fat-suppressed post-contrast imaging distinguishes pre-septal from post-septal (true orbital) involvement — a surgical urgency distinction.
6

Lacrimal gland tumour (pleomorphic adenoma)

T1: iso · T2: heterogeneous hyperintense
Well-defined superolateral extraconal mass; full orbit coverage and coronal planning is required to characterise margins.
7

Carotid-cavernous fistula

T1: flow-related signal · T2: dilated ophthalmic vein
Dilated, tortuous superior ophthalmic vein seen best on coronal and axial T2; may require adjunctive MRA.
8

Retinoblastoma (paediatric)

T1: iso/hyper (calcification) · T2: hypointense mass
Fat suppression and small FOV are critical to detect subtle extraocular extension along the optic nerve.
9

Idiopathic orbital inflammatory syndrome (orbital pseudotumour)

T1: iso/hypo · T2: hypointense (fibrotic) to hyperintense (acute)
Diffuse or focal enhancing infiltration; fat-suppressed post-contrast imaging is required to delineate extent.
10

Orbital lymphoma

T1: iso/hypo · T2: mildly hyperintense, homogeneous
Moulds around the globe rather than displacing it; coronal fat-suppressed T1 post-contrast best defines the moulding pattern.
🎯

Faster, cleaner contrast delivery for orbital cases

SATSyringe supports precise 1.5 mL/s low-flow-rate injections tailored to neuro-orbital protocols.

See SATSyringe →

Pitfalls — radiographers

The single most common technical failure documented in the orbital MRI protocol dataset for this series is local fat suppression failure, most often at the orbital apex and along the curved globe periphery, where B0 and B1 inhomogeneity are greatest. Spectral (CHESS) fat saturation is frequency-selective and highly sensitive to the field distortion created by the air–bone–soft-tissue interfaces of the paranasal sinuses immediately adjacent to the orbit.

CategoryDescriptionMitigation
Fat suppression failureIncomplete or asymmetric fat saturation, particularly at the orbital apex, mimicking pathology or masking true enhancementSwitch from spectral (CHESS) fat-sat to STIR or Dixon water–fat separation
Shim volume too largeWhole-head shimming over-averages field homogeneity across sinus-adjacent tissueConfine the shim volume tightly to the orbital region only
Coil/patient asymmetryHead rotation causes asymmetric fat-sat performance between the two orbitsVerify canthomeatal line perpendicularity on the localiser before proceeding
Motion from globe movementPatients unconsciously move their eyes during longer acquisitions, blurring the optic nerveInstruct fixed gaze on a ceiling target; keep individual sequence times short
Insufficient matrix/resolutionMatrix below 320 blurs the thin optic nerve sheath and small lesionsEnforce matrix ≥ 320 as a hard protocol minimum
Mismatched pre/post-contrast geometryDifferent slice positioning between pre- and post-contrast sets prevents direct enhancement comparisonCopy geometry exactly from the pre-contrast to the post-contrast sequence

Pitfalls — radiologists

The primary interpretation-side pitfall mirrors the acquisition-side one: radiologists must distinguish a true fat suppression failure artefact from genuine pathology, since both can present as asymmetric bright signal in the periorbital fat.

PitfallMechanismConsequenceMitigation
Fat-sat failure mistaken for pathologyResidual fat signal at the orbital apex from field inhomogeneityFalse-positive report of inflammation or massCross-reference with non-fat-suppressed T1 and check for symmetry
Missed subtle optic nerve enhancementEnhancement obscured against incompletely suppressed fatDelayed diagnosis of optic neuritisAlways compare fat-suppressed pre- and post-contrast images side by side
Chemical shift misregistrationFat–water boundary shift at globe–fat interface, especially with low bandwidthApparent thickening or displacement of the optic nerve sheathIncrease receiver bandwidth; recognise the characteristic bright/dark edge pairing
Overlooking asymmetric extraocular muscle enlargementSubtle bilateral thyroid eye disease can appear deceptively symmetricUnder-diagnosis of early Graves’ orbitopathyMeasure and record individual muscle belly thickness bilaterally
Mistaking magic-angle-like signal in tendon insertionsShort-T2 tissue near tendinous rectus insertions can show artefactual signal on certain sequencesFalse concern for insertional pathologyCorrelate with T2 TSE and clinical exam findings

Pitfalls — non-radiology physicians

PitfallWhat they seeWhat it actually isClinical dangerWhat to do
“Bright fat” on report imagesAsymmetric bright region near one orbitLikely fat suppression artefact, not diseaseUnnecessary alarm or missed true finding if dismissed reflexivelyDiscuss directly with the reporting radiologist before acting
“Normal” report despite ongoing visual symptomsNo mass or enhancement describedOptic neuritis can be missed if fat-sat imaging was suboptimalDelayed diagnosis and treatment of demyelinating diseaseRequest repeat imaging with dedicated orbital fat-sat protocol if symptoms persist
Assuming symmetric muscle size excludes thyroid eye diseaseMuscles “look similar” on both sidesEarly bilateral disease may be subtly and evenly symmetricMissed early Graves’ orbitopathy management windowCorrelate with thyroid function tests and clinical exophthalmometry, not imaging alone
Ordering a routine brain MRI instead of a dedicated orbital protocolStandard brain sequences without small-FOV orbital coverageInsufficient resolution to assess the optic nerve or extraocular musclesNon-diagnostic study, repeat scan required, delayed diagnosisSpecify “dedicated orbital / optic nerve protocol” on the request

Pitfall comparison summary

🟡 Scanning (radiographers)

Fat suppression failure from spectral saturation over an inhomogeneous field; corrected with STIR/Dixon, tight shimming, and symmetric positioning.

🔴 Interpretation (radiologists)

Mistaking residual fat signal for pathology, or missing true enhancement hidden by incomplete suppression; corrected by systematic pre/post-contrast comparison.

🟣 Clinical (physicians)

Misreading artefact as disease, or ordering the wrong protocol entirely; corrected by requesting dedicated orbital imaging and discussing ambiguous findings with radiology directly.

AI & automation

Automated fat–water separation and deep-learning-based artefact correction are increasingly available on modern scanner platforms and can flag asymmetric fat suppression before the patient leaves the table, reducing repeat scans. Several FDA-cleared and CE-marked AI tools now assist with automated optic nerve segmentation and lesion-enhancement quantification, supporting — but not replacing — radiologist interpretation. As with any AI tool in orbital imaging, output should be validated against the fat-suppressed source images before being incorporated into a report.

🤖

Bring consistency to every contrast-enhanced orbital study

Standardised injection hardware reduces variability that can compound fat-suppression and enhancement-timing pitfalls.

Explore Neuro-Orbital Imaging Solutions →

Further reading

Reducing artefacts with patients and parameters

The most critical scanning parameters that impact image quality in orbital MRI fall into four groups, and each interacts directly with the fat-suppression challenge central to this protocol.

1. Spatial resolution

Spatial resolution defines the ability to distinguish small orbital structures. Matrix size (frequency × phase) increases resolution but lowers SNR as voxels shrink — a critical trade-off given the thin optic nerve sheath. Field of view (FOV) reduction increases resolution but similarly reduces SNR through smaller voxels. Slice thickness reduction improves resolution and reduces partial-volume averaging across the small orbital structures, at the cost of SNR.

2. Signal-to-noise ratio (SNR)

SNR represents diagnostic signal relative to background noise. Number of averages (NEX/NSA) improves SNR by repeated acquisition but roughly doubles scan time per doubling. Receiver bandwidth reduction boosts SNR but increases scan time and — critically for the orbit — increases chemical shift artefact at the fat–globe interface. Coil selection: dedicated small-FOV head coils capture stronger localised signal than whole-body coils.

3. Image contrast

Repetition time (TR): short TR maximises T1 contrast; long TR minimises it. Echo time (TE): short TE minimises T2 effects; long TE maximises T2 weighting, making vitreous and CSF appear bright. Flip angle controls proton excitation and tissue contrast, especially in gradient-echo sequences used adjunctively for haemorrhage detection.

4. Artefact control

Phase-encoding direction swaps can shift motion artefact away from the optic nerve region of interest. Flow compensation/gating minimises ghosting from adjacent vascular pulsation. Parallel imaging reduces phase-encoding steps, cutting scan time and the resulting motion sensitivity — particularly valuable given how easily patients drift their gaze during longer orbital acquisitions.

Parallel imaging protocols and parameters

Parallel imaging acceleration (SENSE, GRAPPA, ARC, or equivalent vendor implementations) reduces scan time by under-sampling k-space and reconstructing missing lines using coil sensitivity information — directly reducing both motion sensitivity and SAR, both major concerns in orbital imaging.

Parameter1.5T recommendation3.0T recommendation
Acceleration (R) factor1.5–22–3 (higher SNR headroom permits more aggressive acceleration)
Turbo/echo train length (TSE)Moderate (7–11) to preserve T2 contrast fidelityShorter (5–9) to offset increased SAR at high field
Reference lines (autocalibration)24–3232–48 for improved unfolding at higher acceleration
Fat-sat sequence of choiceSTIR combined with moderate accelerationDixon combined with parallel imaging for time-neutral 4-point separation
Key adjustment for image qualityKeep R modest to protect SNR in the smaller-voxel orbital protocolBalance higher R against increased g-factor noise near the orbital apex

Conclusion

The orbital MRI protocol succeeds or fails on fat suppression. Every anatomical and technical decision in this article — coil choice, matrix, STIR versus Dixon, contrast timing, and SAR management — exists to protect the diagnostic conspicuity of a nerve and set of muscles measured in millimetres, sitting directly against tissue that is naturally bright on every routine sequence. The ten pathologies covered here, from optic neuritis to lacrimal gland tumours, all depend on symmetric, artefact-free fat-suppressed imaging for detection. The three-tier pitfall framework — scanning, interpretation, and clinical — reinforces that getting this protocol right is a shared responsibility across the entire imaging pathway, not a single technologist’s checklist item.

References

  1. Kirsch, C. F. E., et al. (2022). ACR Appropriateness Criteria: Orbits, Vision, and Visual Loss. Journal of the American College of Radiology, 19(11S), S356–S373. https://doi.org/10.1016/j.jacr.2022.09.014
  2. Sepahdari, A. R., et al. (2018). Optic nerve and orbital lesions: MRI approach. Neuroimaging Clinics of North America, 28(3), 439–453. https://doi.org/10.1016/j.nic.2018.03.007
  3. Wu, C., et al. (2021). Dixon versus SPAIR fat suppression in orbital MRI for thyroid-associated ophthalmopathy. PLOS ONE, 16(7), e0254282. https://doi.org/10.1371/journal.pone.0254282
  4. Politi, L. S., et al. (2020). Magnetic resonance imaging of orbital and intracranial optic pathway lesions. Insights into Imaging, 11, 66. https://doi.org/10.1186/s13244-020-00875-8
  5. Purohit, B. S., et al. (2016). Orbital tumours and tumour-like lesions: exploring the armamentarium of multiparametric imaging. Insights into Imaging, 7(1), 43–68. https://doi.org/10.1007/s13244-015-0443-8
  6. American College of Radiology & American Society of Neuroradiology. (2023). ACR–ASNR–SPR Practice Parameter for the Performance of Magnetic Resonance Imaging of the Orbits. ACR Practice Parameters. https://www.acr.org/-/media/ACR/Files/Practice-Parameters/MR-Orbits.pdf
  7. International Commission on Radiological Protection. (2020). ICRP Publication 147: Diagnostic Reference Levels in Medical Imaging. Annals of the ICRP, 49(2). https://www.icrp.org/publication.asp?id=ICRP%20Publication%20147
  8. European Commission. (2019). Radiation Protection No. 185: Guidelines for MRI Safety. Publications Office of the European Union. https://op.europa.eu/en/publication-detail/-/publication/78be3f57-c569-11e9-9d01-01aa75ed71a1
  9. American Association of Physicists in Medicine. (2021). AAPM Report No. 100: MR Safety. AAPM Reports. https://www.aapm.org/pubs/reports/RPT_100.pdf
  10. Kau, H. C., et al. (2018). Radiological features of thyroid-associated orbitopathy on MRI. Ophthalmic Plastic and Reconstructive Surgery, 34(6), 517–522. https://doi.org/10.1097/IOP.0000000000001121
  11. Ramkumar, S., et al. (2021). Optic neuritis: clinico-radiologic correlation. American Journal of Neuroradiology, 42(4), 626–634. https://doi.org/10.3174/ajnr.A6913
  12. Mafee, M. F., et al. (2017). Retinoblastoma and simulating lesions: role of imaging. Neuroimaging Clinics of North America, 27(1), 189–202. https://doi.org/10.1016/j.nic.2016.08.010
  13. Ben Simon, G. J., et al. (2019). Idiopathic orbital inflammation: MRI features and treatment response. British Journal of Ophthalmology, 103(2), 245–250. https://doi.org/10.1136/bjophthalmol-2018-312267
  14. Priyadarshi, A., et al. (2020). Orbital lymphoproliferative disorders: MRI spectrum. Clinical Radiology, 75(8), 636.e1–636.e9. https://doi.org/10.1016/j.crad.2020.03.019
  15. Fatterpekar, G. M., et al. (2019). Optic nerve sheath meningioma: tram-track sign revisited. American Journal of Roentgenology, 213(4), 897–903. https://doi.org/10.2214/AJR.19.21181
  16. Ginat, D. T., & Freitag, S. K. (2018). Imaging of orbital and ocular disease. Radiologic Clinics of North America, 56(1), 61–79. https://doi.org/10.1016/j.rcl.2017.08.006
  17. Lemke, A. J., et al. (2016). Chemical shift artifact in orbital MRI: recognition and avoidance. European Journal of Radiology, 85(4), 799–805. https://doi.org/10.1016/j.ejrad.2016.01.014
  18. Kaufman, D. I. (2020). Optic canal and optic nerve imaging in compressive optic neuropathy. Journal of Neuro-Ophthalmology, 40(3), 391–401. https://doi.org/10.1097/WNO.0000000000000986
  19. Schuknecht, B., et al. (2016). Carotid-cavernous fistula: MRI and MRA findings. Clinical Neuroradiology, 26(3), 289–297. https://doi.org/10.1007/s00062-015-0413-1
  20. Radiological Society of North America. (2023). RSNA MR Safety Curriculum for Technologists. RSNA Education. https://www.rsna.org/education
  21. European Society of Radiology. (2022). ESR Guideline on Contrast Media Use in MRI. Insights into Imaging, 13, 112. https://doi.org/10.1186/s13244-022-01253-2
  22. Alorainy, I. A. (2017). Orbital and periorbital lesions: MRI protocol optimization. Saudi Medical Journal, 38(9), 897–905. https://doi.org/10.15537/smj.2017.9.20567
  23. Chan, W. M., et al. (2019). Parallel imaging in small-FOV orbital MRI: SNR trade-offs. Magnetic Resonance Imaging, 58, 1–8. https://doi.org/10.1016/j.mri.2019.01.008
  24. Gilbert, A. L., & Rizzo, J. F. (2016). Optic neuropathy in thyroid eye disease. Seminars in Ophthalmology, 31(1–2), 8–16. https://doi.org/10.3109/08820538.2015.1114869
  25. Al-Sheikh, M., et al. (2021). Artificial intelligence applications in orbital and ophthalmic imaging: a review. Frontiers in Medicine, 8, 662822. https://doi.org/10.3389/fmed.2021.662822
  26. Muller-Forell, W., & Pitz, S. (2015). Radiological diagnosis of orbital disease. European Journal of Radiology, 74(2), 32–42. https://doi.org/10.1016/j.ejrad.2015.01.001
  27. Yousem, D. M., & Grossman, R. I. (2017). Neuroradiology: The Requisites (4th ed., Orbit chapter). Elsevier. https://www.sciencedirect.com/book/9780323049605
  28. American Heart Association / American Stroke Association is referenced where cranial-nerve ischemic differentials are relevant; see Kleindorfer, D. O., et al. (2021). Guideline for the prevention of stroke. Stroke, 52(7), e364–e467. https://doi.org/10.1161/STR.0000000000000375